ANALYZER 300 Specifications:
 
Specifications subject to change without notice.

MEASUREMENT SETUP AND PRESENTATION
Display The display is divided into six user defined rows and three or five columns. For each row, a suitable parameter and measurement mode may be selected
Parameters
Voltage (U)
Power factor (PF)
Current (I) Phase angle (phi)
Frequency (f) Active energy (W)
Active power (P) Reactive energy (Wr)
Reactive power (Q) Apparent energy (Wa)
Apparent power (S) Crest factor (CF)
THD (I and U) Total Harmonic Distortion (%)
Measurement mode AC, DC or AC+DC
Frequency range DC and 40 - 70Hz fundamental
Frequency source Voltage or current input
Measurement window 0.5 - 5 seconds
Averaging Moving of 1 - 100 measurements

VOLTAGE - U
Ranges 2000/600/200/60/20/6 V peak or DC
Maximum input voltage U+to U- or U+/U- to I+/I- 600Vrms (AC+DC) or 2000Vpeak. U+/U- to case 400Vrms (AC+DC)
Uncertainty,
in rms, at 23±5°
AC
DC,AC+DC
Peak
±(0.2% of rdg + 0.05% of range)
±(AC spec. + 15mV)
±(AC spec. + 0.1% of rdg)
Display resolution Better than 0.01% of range or 1mV
Input impedance 2x1M//470pF

CURRENT - I
Ranges (AC peak and DC) 300A / 100A / 30A / 10A / 3A / 1A / 0.3A / 0.1A Automatic or manual range selection
Maximum input current 20Arms (AC+DC) continuous 300Apeak or 200Arms 20ms every 2sec
Uncertainty,
in rms, at 23±5°
AC
DC,AC+DC
Peak
±(0.3% of rdg + 0.05% of range)
±(AC spec. + 0.5mA)
±(AC spec + 0.1% of rdg)
Display resolution Better than 0.02% of range or 10µA
Input impedance ~18mOhm

POWER - P,Q and S
Ranges 48 ranges from 0.6VA to 12kVA. Automatic or manual selection of voltage and current range
Maximum input 600Vrms and 20Arms
Uncertainty at 23±5° P + Q AC
S
 
±(0.4% of rdg + 0.1% of range)
±(0.4+D)% of rdg S + 0.1% of range
D=0.01*(UTHD%)*(ITHD%)
Display resolution Better than 0.02% of range or 0.1mVA

Energy - W, Wr and Wa
Measurement range 0 - 999 999 kWh
Timer uncertainty ±0.01% (fixed measurement window)
Display resolution Better than 0.05% of rdg or 0.02% of (power range *1h)

Charge - Ch
Measurement range 0 - 999 999 Ah
Timer uncertainty ±0.01% (fixed measurement window)
Display resolution Better than 0.05% of rdg or 0.02% of (power range *1h)

Frequency - F
Measurement range 40 - 70Hz
Source Voltage or current input
Principle Auto correlation
Uncertainty ±0.01% of reading
Resolution 0.001Hz

GENERAL
Compliance meas. acc. to IEC555-2, EN60555-2, IEC1000-3-2 and EN 61000-3-2
Testing Pre-programmed limits according to standards for pass/fail testing. User specified limits may be added
External power source Remote control via GPIB. Selection of voltage and frequency from ANALYZER 300

PRESENTATION
Display Selectable between table and graphical presentation of harmonic rms values
Limits On or off, both in graph and table
Displayed parameters Total rms value of U and I, source frequency, active power, rms value of fundamental and THD

MEASUREMENT
Harmonic order 1 - 40
Frequency of fundamental 40 - 70Hz
Frequency source Voltage or current
Data source Voltage or current
Calculation method FFT with 32 bit floating point math
AD-conversion 18 bits resolution
Measurement window Rectangular
Analyzed periods 1- 47
Synchronization uncertainty ±0.01%
Antialiasing filter attenuation 80dB
Measurement time DSP
<1 second when analyzing 6 periods
Uncert. at 23±5°C, rms ±(0.5% of rdg + 0.03% of range)

GENERAL
Compliance meas. acc. to IEC555-3, EN60555-3, IEC1000-3-3 and EN 61000-3-2
Method of analysis Implementation of IEC868/EN60868, flicker meter specification
Testing Pre-programmed limits according to standard for pass/fail testing. User specified limits may be added
External power source Remote control via GPIB. Selection of voltage and frequency from ANALYZER 300

PRESENTATION
Displayed parameters
Pst Short-term flicker
Plt Long-term flicker
dc Relative steady state voltage change
dmax Max relative voltage change
d(t) Relative voltage change
Test voltage and frequency with deviations for the collection period
Limits On or off for pass/fail decision

MEASUREMENT
Range Pst
  Plt
  d
0.1 to 20
0.2 to 20
0 to 25%
Frequency range 40 - 70Hz
Reference impedance R + jX simulated in calculation with compensation possibility for physical reference impedance
Measurement time 1 - 15 minutes
Number of measurements 1 - 1100
Uncertainty at 23±5° Pst
Plt
d
± 4% of rdg. for 0.5 < Pst < 20
± 4% of rdg. for 0.5 < Plt < 20
± 2% of rdg. for dmax > 0.1%

GENERAL
Display Simultaneous recording of up to three user defined parameters
Magnification Any trace may be selected to use full screen
Memory Freeze of recording may be used any time without loosing information
Number of traces 1 - 3

SELECTIONS FOR EACH TRACE
Channel M1, M2 or M3
Parameter
Voltage (U) Frequency (f)
Current (I) Crest factor (CF)
Active power (P) Power factor (PF)
Reactive power (Q) Phase angle (phi)
Apparent power (S)  
Measurement mode AC, DC or AC+DC

MEASUREMENT
Frequency range DC and 40 - 70Hz
Frequency source Voltage or current
Horizontal axis length 8, 15, or 30 min, 1, 2, 4, 8 or 24 hours
Measurements per incr. 1,2,3,8,15,30,60,180
Measurement window 1 s

GENERAL
Display Simultaneous presentation of waveforms. At single phase, voltage and current. At three phase, voltage or current for all phases
Testing Mask for pass/fail testing acc. to IEC1000-3-2 class D
Wave measurements A vertical cursor is available for on-wave measurements

PRESENTATION
Displayed parameters Rms values of voltage and current. Frequency of frequency source
Cursor parameters Horizontal position (phi), vertical position (u), relative time (dt) and relative frequency (df)
No. of waveforms 2 at single phase and 3 at three phase
Magnification Any wave may be zoomed to full screen

MEASUREMENT
Frequency range 40 - 70Hz fundamental
Frequency source Voltage or current
Measurement mode AC+DC
Horizontal axis length 1,2,4,8 or 16 periods
Vertical axis Normalized to peak values

Applications Up to 4 individual instrument set-ups may be stored as applications in an internal non volatile memory.More applications may be stored and read from disk
Data Collected data from recording, waveform or harmonics measurements may be stored and read from disk
Data format dBase DBF

Interface GPIB (IEEE 488.1-87) or RS-232
Connectors GPIB
  RS-232
According to IEEE 488.1-87
25 pin DSUB
GPIB interface capabilities Talker, Listener and Controller. SH1, T6, AH1, L4, SR1, RL1, PP0, DC1, DT1, (C0), E2
Remote control language IEEE 488.2-87 and SCPI-1994.0
GPIB address User selectable to 0 - 30

Display LCD 640x480 resolution with backlight
Printer output Parallel, Centronics compatible
Floppy drive 1.44MB 3½" PC-format
Rack mounting With optional rack mounting kit. Size 19" 3HE
Dimensions (HxWxL) 132x425x340mm (5.2x16.7x13.4 inches)
Weight Single phase 9kg (20 lbs), three phase 11.4kg (25 lbs)
Operating environment 0 to +40° <80% R.H. non-condensing
Storage environment -30 to +60° non-condensing
Power supply 100-130V or 200-260V, automatic range selection
Power line frequency 45 - 65 Hz
Power consumption 45 W max
Fuses Instrument power 2xF1A, Current input Littlefuse KLK25A
CE complliance EN50081-1:92 and EN50082-1:92 ,EN61010-1:93. Installation Category II, CAT II

 

    This page was updated 2008-02-05